GREEN  SAFETY  RELIABLE  EFFICIENT
24-hour hot line:+86 189 1133 7382
sales@auwii.com
​​Committed to laboratory sample pretreatment and safety protection equipment R&D and manufacturing
How does Auwii Acid Evaporator battle Result in Korean Food lab ?
Source: | Author:AUWII | time :2020-12-31 | 719 Browse: | Share:
Test Purpose : To evaporate acid from a sample treated with MDS in order to reduce the nitric acid concentration of the final diluted sample to less than 10%.

How does Auwii Acid Evaporator battle Result in Korean Food lab

1.   Customer :DS Foods Corporation, R&D Center

2.   Analytical Instrument : PerkinElmer ICP-OES OPTIMA 8300DV

3.  Test Purpose : To evaporate acid from a sample treated with MDS in order to reduce the nitric acid concentration of the final diluted sample to less than 10%.

 (Follow ICP-OES test method for food analysis by Korea FDA)

[Test Result of Acid Evaporator]

1) Recovery Test for Cd

* Certified Value: 1.6 ppm [Cd]

* Sample: Rice flour(Certified Reference Material)

 2) Analysis Progress

- Target Sample Volume : 18~20 of MDS Digested sample was evaporated to less than 1 .

- Setting Temp. : 180

- Time : Reach to set value : 40~60 min

           Evaporation : 60 min

           Total Time : 2 hrs

à Notes : Microwave Digested samples amount of 8 ~ 9 ml and additional DI Water amount for washing the outer wall of the MDS vessel combined became C. 18 to 20 ml of total evaporation fluid.

à Evaluation : The temperature controller is attached to the main body of the machine, making it easy to use.

Additional Comment : The recovery acceptable range is 95 to 110 %.  Most of them are within the acceptable range and are satisfactory.

“Compared to a competitor’s equipment, it is easier to use and integrated design saves space.”

Users comment like that.

Congratulations to Auwii!